The electrochemical chlorine sensors from WTW for measurements in drinking water or swimming pools are connected with the ADA Cl/IDS to the MIQ/IDS module (FMCL 412 M12) or directly to the CL 298 converter (FCML 412 N or TCML N):
With a uniform measuring range, the sensors must be selected depending on the pH value of the applications.
FCML 412 N and TCML N: The chlorine electrodes for free or total chlorine already have a 2-wire connection cable and can therefore be quickly and easily connected to the CL 298 converter.
FMCL 412 M12: The chlorine electrodes require the ADA Cl/IDS adapter cable in order to be connected to the IQ Sensor Net with the MIQ/IDS module. This allows them to be combined as a digital sensor for continuous monitoring with the IQ SN controllers DIQ/S 28x or MIQ/TC 2020 3G.
FCML 412 M12-2 - Electrode for free chlorine up to 2 mg/l - Connection to MIQ/IDS
FCML 412 M12-20 - Electrode for free chlorine up to 20 mg/l - Connection to MIQ/IDS
ADA Cl/IDS - Adapter for connecting FCML 412-M12 to MIQ/IDS modules
FCML 412 N - Electrode for free chlorine - Analog connection to Cl 298
TCML 412 N - Electrode for total chlorine - Analog connection to Cl 298
Model | FCML 412 N | FCML 412-M12-2 | FCML 412-M12-20 | TCML N |
Order No. | 201187 | 201189 | 201194 | 201192 |
detailed technical data see data sheet D7.01 | ||||
Measuring principle | Amperometric | |||
Measured value | Free chlorine | Total chlorine | ||
Measuring range | 0.01 … 2.00 mg/l Cl2 | 0.005 ... 2.000 mg/l Cl2 | 0.05 ... 20.00 mg/l Cl2 | 0.01 … 2.00 mg/l Cl2 |
Response time | t90 approx. 120 s | |||
Minimum flow rate | Recommended minimum flow rate in flow cell D 19: 15 l/h, or alternatively in flow cell D-CL: > 30 l/h | |||
Temperature measurement | 0 … 45 °C | |||
Temperature compensation | Automatically via integrated sensor | |||
pH range | 4 … 9 | 4 … 12 |