Validation logger for benchtop autoclaves, 1340-6648, EBI 12-TP460 Ebro Germany
  • Validation logger for benchtop autoclaves, 1340-6648, EBI 12-TP460 Ebro Germany

Validation logger for benchtop autoclaves, 1340-6648, EBI 12-TP460 Ebro Germany

Price: 0৳
Status : Call For Price
Brand: Ebro Germany
  • Model: EBI 12-TP460
  • Product Code: 971
  • Availability: Call For Price
Key Features
  • Bendable sensor for placement in cavities
  • Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning. With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers. The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time. When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.
  • View More Info
+8801325065151
Call For Price

Description :

Validation logger for benchtop autoclaves, 1340-6648, EBI 12-TP460 Ebro Germany

 

bendable sensor for placement in cavities

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning. With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers. The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time. When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.

Benefits:

  • highly flexible cable probes
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger


Applications:

  • Validation logger for Measurement in the steam sterilizer

 

Specification :

 

Data memory

4 x 20,064 measurement values

Weight

aprox.120 g

Accuracy

±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)

±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),  
± 0,5 % FS (3.500 mbar ... 4.000 mbar)

Protection class

IP68

Dimensions

ø 48 mm x 24 mm - Needle length 600 mm

Resolution

0.01 °C 1 mbar

Certificate

Factory calibration certificate

Measured variables

Temperature, pressure

Intervall

250 ms ... 24 h

Battery

3.6 V lithium battery exchangeable

Measuring range

Temperature 0 °C … +150 °C Pressure 1 … 4.000 mbar

Storage temperature

-20 °C ... +70 °C

Reviews (0) :

Get specific details about this product from customers who own it.

Write your review

 Bad           Good