Validation logger to measure in steam sterilizer, 1340-6640, EBI 12-T470 Ebro Germany
bendable
sensor for placement in cavities
Validation logger for
thermal testing of processes with steam sterilizers. The logger is
characterized by its highly flexible cable probes, which enable precise
positioning. When using the logger in the WD, it is strongly recommended to fix
the probes by means of silicone cable ties. The very thin temperature sensors
with a diameter of 1.2 mm at the sensor tip enable a very fast response time.
Benefits:
Applications:
Specification :
Data memory
2 x 50,160
measurement values
Sensor
Pt 1,000,
class A
Weight
approx. 110 g
Accuracy
± 0,2 °C (-20
°C … 0 °C) ± 0,1 °C (0 °C ... +120 °C) ± 0,05 °C (+120 °C … +140 °C) ± 0,1 °C
(+140 °C … +150 °C)
Protection
class
IP68
Dimensions
(Ø x H) 48 mm
x 24 mm - Needle length 600 mm
Resolution
0,01 °C
Certificate
Factory
Calibration: 0 °C, +60 °C and +134 °C
Measured
variables
Temperature
Intervall
250 msec ...
24 h
Battery
Lithium
battery (3,6 V), user replaceable
Measuring
range
-20 °C … +150
°C
Storage
temperature
-20 °C ... +70 °C